Draft Details
Designation:CSA ISO 15883 -7
Source:CSA
Contact:[email protected]
Contact name:Wesley Salifu
Review start date:Apr 24, 2026
Review end date:Jun 23, 2026
Categories:Health Care and Medical Devices
Contact email:angelique.coyle(at)csagroup.org
Associated materials (click link to download). You will need to be logged in to download any files:
- ISO 15883-7_2025_PR.pdf (PDF document, 1 MB)
For PDF files, Adobe Acrobat Reader can be downloaded from the Adobe website.
Draft Scope/Description:
This document specifies the requirements for washer-disinfectors (WD) intended to be used for the cleaning and chemical disinfection, in a single operating cycle, of reusable items such as:
a) bed frames;
b) bedside tables;
c) transport carts;
d) containers;
e) surgical tables;
f) sterilization containers;
g) surgical clogs;
h) wheelchairs;
i) aids for persons with disabilities.
This document also specifies the performance requirements for the cleaning and disinfection of the WD and its components and accessories.
Devices identified within the scopes of ISO 15883-2, ISO 15883-3, ISO 15883-4, and ISO 15883-6 do not fall within the scope of this document.
In addition, this document specifies the methods for type testing, works testing, validation (installation, operation, and performance qualification on first installation), routine control, and monitoring, as well as requalifications to be carried out periodically and after essential repairs.
NOTE 1 WD covered by this document can also be used for cleaning and chemical disinfection of other thermolabile and reusable devices as recommended in the instructions for use (IFU) for those devices.
NOTE 2 The performance requirements specified in this document cannot ensure the inactivation or removal of the causative agent(s) (prion proteins) of transmissible spongiform encephalopathies.
You may comment on any section of this document by clicking the “Submit Comment” link at the bottom of the relevant section.