Draft Details
Designation:CSA Z15883-2
Source:CSA
Contact:[email protected]
Contact name:Wesley Salifu
Review start date:Apr 24, 2026
Review end date:Jun 23, 2026
Categories:Health Care and Medical Devices
Contact email:angelique.coyle(at)csagroup.org
Associated materials (click link to download). You will need to be logged in to download any files:
- ISO 15883-2_2024_PR.pdf (PDF document, 1 MB)
For PDF files, Adobe Acrobat Reader can be downloaded from the Adobe website.
Draft Scope/Description:
This document specifies requirements for washer-disinfectors (WD) that are intended for use for the cleaning and thermal disinfection, in a single operating cycle, of reusable critical and semi-critical medical devices, such as surgical instruments, anaesthetic equipment, and any non-critical devices used in conjunction with critical and semi-critical medical devices, such as bowls, dishes and receivers, utensils and glassware.
This document is intended to be used in conjunction with the general requirements specified in ISO 15883-1:2024, except those specified in 4.1.1.
NOTE The specified performance requirements of this document cannot ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.
You may comment on any section of this document by clicking the “Submit Comment” link at the bottom of the relevant section.