201.1.1 *Scope
[Add the following paragraph]
This Standard applies to ME EQUIPMENT and ME SYSTEMS that are intended to be installed or used in accordance with CSA C22.1, Canadian Electrical Code, Part I.
IEC Scope
Clause 1 of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 applies, except as follows:
201.1.1 *Scope
Replacement:
This part of IEC 60601 applies to BASIC SAFETY and ESSENTIAL PERFORMANCE of INVASIVE BLOOD PRESSURE MONITORING EQUIPMENT as defined in 201.3.63, hereinafter also referred to as ME EQUIPMENT.
This document applies to INVASIVE BLOOD PRESSURE MONITORING EQUIPMENT intended for use in professional healthcare facilities and in the EMERGENCY MEDICAL SERVICE ENVIRONMENT.
This document does not apply to catheter tubing, catheter needles, Luer locks, taps and tap tables that connect to the DOME.
This document does not apply to non-invasive blood pressure monitoring equipment.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as follows:
The clause or subclause applies to ME EQUIPMENT, as default and, only if the corresponding safety measure or function is not completely integrated into the ME EQUIPMENT but implemented as part of an ME SYSTEM, the clause or subclause applies to the ME SYSTEM.