Draft Scope/Description:

This second edition cancels and replaces the first edition published in 2009 and amendment 1:2013. This new edition constitutes a technical revisions 

This edition includes the following significant technical changes with respect to the 

previous edition: 
- Reference revision and Alignment with 
o IEC 60601-1:2005 + AMD1:2012 
o IEC 60601-1-8:2006 + AMD1:2012 
o IEC 60601-1-10:2007 
o IEC 60601-1-12 
- Changing an operator-accessible cuff-sphygmomanometer connector from not 
compatible with the ISO 594 series to compatible with the ISO 80369 series 
- Added additional requirements for public self-use sphygmomanometers 
- Added a list of primary operating functions. 

 